Products & Solutions

Products and Solutions

pressure

Digital pressure gauge ACD-201
Digital pressure gauge ACD-201
Digital pressure gauge ACD-201
Digital pressure gauge ACD-201

Digital pressure gauge ACD-201

The ACD-201 remote digital pressure gauge adds a communication module for remote output on the basis of displaying the digital pressure gauge locally. It has RS485 communication, USB communication, (4~20) mA current signal and other output methods. With communication software, it can directly communicate with the computer to save, process and report detection data; Adopting a fully isolated signal design to eliminate interference; Built in lithium battery power supply, with a lifespan of 5-8 years, can still stably collect pressure even when the external power supply is disconnected (communication function failure). The appearance is available in radial, axial, and disc mounting options.

Main Features

display unit

Pa , kPa , MPa , psi , bar , mbar , kgf/cm2 , inH2O , mmH2O, inHg, mmHg

measurement range

gauge pressure  :-0.1MPa~0~...~260MPa

differential pressure :0~500Pa~...~3.5MPa

Absolute pressure :0~0.035MPa~...~60MPa

accuracy class

0.05 level, 0.1 level, 0.2 level, 0.5 level

output signal  

(4-20) mA (24V DC, two-wire system), (1-5) V (24V DC, three wire system)

contact method    

RS485 、USB

range transfer      

Suggest not exceeding 3:1 (for current output signals)

load capacity

RL  = U-12(V) / 0.02(A)  

U--Loop voltage(V)  

RL--Allow load resistance(Ω)

power supply method  

Internal power: Built in 1 3.6V high-performance lithium battery

External power: (10~30) V DC

Collection speed

(0.25~10) seconds/time, default 0.5 seconds/time, time can be set

stable performance

Better than 0.1% FS/year

Battery working time

Built in 3.6V high-performance lithium battery, reference conditions: (16-25) ℃

image

ambient temperature

 (-30~70)℃  

relative humidity

<90%

atmospheric pressure

(86~106)kPa

other

Calibration reference environment temperature: 20 ℃± 2 ℃
0.05 Level instrument working environment temperature: (0-50) ℃ 

Measure the temperature of the medium

routine  (-40~120)℃

wide temperature  (-60~150)℃

display mode

Five digit display+percentage bar chart

Protection level

IP65

Explosion-proof rating

ExiaIICT4 Ga  

Overload pressure

1.5 to 3 times the measuring range, depending on the measurement range

software configuration

AncnView-T intelligent instrument 

detection software (with RS485, USB communication)

 displays instrument data in real time,

utomatically records and draws pressure curves,

and can be exported to an Excel 

spreadsheet for reading, printing, and saving.

process connection

Public  Manufacturing: M20 × 1.5, M14 × 1.5, M14 × 1, etc.
English/American standards: G1/2, G3/4, G1/8, G3/8
NPT 1/2, NPT 1/4, NPT 1/8, etc.

 